Bump 30 – RKI Instruments GX-2009 Manual User Manual

Page 46

Advertising
background image

41 • Calibration Mode

GX-2009 Operator’s Manual

NOTE:

The bump test gas concentrations are the same as the AUTO
CAL gas concentrations.

If the values are not correct, you can change them by changing the
AUTO CAL gas values in Calibration Mode or User Setup Mode or by
using the Data Logger Management Program or the User Setup
Program. To exit this screen and return to the main menu, press and
release the AIR button.

3. Use the sample tubing to connect the calibration adapter plate to the

regulator. Attach the tubing to the adapter plate on the inlet side as
shown below in Figure 13.

Figure 13: Calibration Kit Assembly

4. Confirm that the regulator on/off knob is turned all the way clockwise

(closed) and screw the calibration gas cylinder onto the regulator.

50 12.0

BUMP 30

CO ppm H2S ppm

CH4 %LEL OXY vol%

50 25.0

O

2

CO

Flow

To Fixed Flow

Regulator

Calibration Tubing

T

T

u b

in g

Adapter Plate

%L

E

L

H

2

S