23 • Operation

GX-2009 Operator’s Manual

Alarm Logic/Resetting Alarms

The gas alarm logic can be set to either latching or self resetting with the
GX-2009 Setup Program by setting the Alarm Pattern parameter. See the
GX-2009 Setup Program Operator’s Manual for information on how to set
the alarm logic. If Alarm Pattern is set to Latch, the gas alarms are
latching. If Alarm Pattern is set to Reset, the gas alarms are self-resetting.
The standard factory setting is Latch for latching alarms.

If the alarms are latching, then an alarm indication will continue even if the
gas reading causing the alarm decreases below the alarm setpoint
(increases above for oxygen Warning) and will continue until the alarm is
reset. To be able to reset the alarm, the gas reading that caused the alarm
must decrease below the alarm setpoint (increase above for oxygen
Warning) before that alarm can be reset. To reset latching alarms, press
and release the POWER MODE button after the gas reading falls below the

Over Range
Concentration of gas rises
above full scale

OVER flashes to the

right of the battery level
icon at the same rate as
Alarm indication

• Gas reading replaced by

brackets flashing at the
same rate as Alarm
indication

• Alarm LEDs flash at the

same rate as Alarm
indication

• Backlight turns on

• Buzzer sounds

alternating between
a low and high pitch
at same rate as
Alarm indication

• Vibrator pulses at

same rate as Alarm
indication

Low Battery Warning

• Battery icon blinks

None

Dead Battery Alarm

• Gas readings replaced

by FAIL in the upper left
corner of LCD

BATTERY is displayed

to the right of the battery
icon.

• Battery icon blinks
• LEDs flash

Buzzer pulses

Sensor Failure

• Name of failed sensor or

sensors displayed.

FAIL is displayed
• LEDs flash

Buzzer sounds a
double pulsing tone
(two pulses in quick
succession)

Table 4: Alarm Types and Indications

Alarm Type

Visual Indications

Other Indications