63 • Maintenance

GX-2009 Operator’s Manual

7. Reattach the sensor cover retainer in its original position. It should snap

into place onto the sensor retainer.

8. Replace the two screws that you removed in Step 2 above to secure the

sensor retainer and sensor cover retainer to the GX-2009.

Replacing the Sensor Cover

WARNING: Replace the sensor cover in a non-hazardous

environment.

1. Verify that the GX-2009 is off.

2. Unscrew the two screws that secure the sensor retainer and sensor

cover retainer to the GX-2009 and remove them.

3. Using a small flat blade screwdriver, gently pry each side of the sensor

cover retainer away from the sensor retainer.

4. Remove the sensor cover then insert the replacement cover.

5. Reattach the sensor cover retainer in its original position. It should snap

into place onto the sensor retainer.

6. Replace the two screws that you removed in Step 2 above to secure the

sensor retainer and sensor cover retainer to the GX-2009.

Replacing a Sensor

WARNING: Replace the sensors in a non-hazardous environment.

1. Verify that the GX-2009 is off.

2. Unscrew the two screws that secure the sensor retainer and the sensor

cover retainer to the GX-2009 and remove them.

3. Use a small flat blade screwdriver to pry each end the sensor retainer

from the case with the sensor cover retainer still attached.

Advertising