GX-2009 Operator’s Manual

Calibration Mode • 34

3. Use the sample tubing to connect the calibration adapter plate to the

regulator. Attach the tubing to the adapter plate on the inlet side as
shown below in Figure 9.

Figure 9: Calibration Kit Assembly

4. Confirm that the regulator on/off knob is turned all the way clockwise

(closed) and screw the calibration gas cylinder onto the regulator.

5. Push the adapter plate onto the GX-2009’s sensor face. Make sure the

adapter plate is oriented as shown in Figure 10 below with the sensor
names on the adapter plate matching up with the sensor names on the
instrument.

Figure 10: Installing the Adapter Plate, Auto Cal

O

2

CO

Flow

To Fixed Flow

Regulator

Calibration Tubing

T

T

u b

in g

Adapter Plate

%L
E

L

H

2

S

Flow

%L
E

L

H

2

S

O

2

To Fixed Flow

Regulator

CO

AIR

POWER

MODE

RKI GX-2009

H

2

S CO %LEL O

2

Adapter Plate