15 • Start Up

GX-2009 Operator’s Manual

WARNING:You must perform a bump test in order to continue to

normal operation. If you do not perform a bump test, the
buzzer will continue to beep and the LEDs will continue to
flash for 6 seconds every 5 seconds and the unit will not
enter normal operation.

If bump testing is not due or Bump Test Limit Check is set to None,
then the following screen appears for a few seconds indicating when
the next bump test is due.

NOTE: If a successful calibration is performed, the next bump test date is

reset and starts over even though a bump test was not performed.

5. The Date/Time Screen appears for a few seconds.

This screen displays the current date and time.

6. The Battery Voltage/Alarm Pattern Screen appears for a few seconds.

The screen displays the current battery voltage and the alarm pattern.
AL -- H indicates latching alarms and AL -- A indicates self-resetting
alarms. See “Alarm Logic/Resetting Alarms” on page 23 for a
description of latching and self-resetting alarms.

7. The display then indicates the following items for about a second each:

Full scale values for all channels

Warning setpoint (low gas alarm) for all channels

Alarm setpoint (high gas alarm) for all channels

STEL alarm setpoint for the CO and H2S channels

TWA alarm setpoint for the CO and H2S channels

08 - 7.27

NEXT BUMP

15 d

08 - 7.27

3:51

bAtt.

2.4

V

AL -- H