51 • User Setup Mode

GX-2009 Operator’s Manual

3. When the desired channel is displayed, press and release the POWER

MODE button to select the channel for an alarm point adjustment. In the
example below, the combustible channel has been selected and the
Warning alarm point is displayed flashing.

4. Use the AIR button to adjust the Warning setpoint to the desired value.

5. Press and release the POWER MODE button to save the new Warning

setpoint. A screen with the Alarm setpoint flashing is displayed.

6. Use the AIR button to adjust the Alarm setpoint to the desired value.

7. Press and release the POWER MODE button to save the new Alarm

setpoint. The display will return to the following screen.

NOTE:

If you are setting the H2S or CO channel’s alarm points, you will
also see screens for setting the TWA and STEL alarms.

8. Repeat Step 2 through Step 7 for any other channels whose alarm

points you want to change.

9. When you are done, scroll to the following screen.

10. Press and release the POWER MODE button to return to the

ALARM--P screen.

10

WARNING

CH4 %LEL

50

ALARM

CH4 %LEL

- - -

ALARM--P

CH4 %LEL

ESCAPE

ALARM--P