GX-2009 Operator’s Manual

User Setup Mode • 50

Setting the Alarm Points

Entering the ALARM--P menu item allows you to set the alarm points for
each channel.

Table 5 below lists the factory set alarm points for each channel.

All of the alarms are increasing alarms, the alarm condition actuates when
the gas level increases above the alarm point, except for the O

2

Warning

alarm. The O

2

Warning alarm is a decreasing alarm. The O

2

Warning

condition actuates when the oxygen level decreases below the alarm point.

For the combustibles channel, it is not possible to adjust either the Warning
or Alarm settings to be higher than 60 %LEL. In addition, the Warning
setting cannot be adjusted higher than the Alarm setting for all channels.

1. At the ALARM--P screen, press and release the POWER MODE button.

The following screen will display showing the combustible channel.

2. You can scroll through all the channels by pressing and releasing the

AIR button.

Table 5: Factory Set Alarm Points

Channel

Warning

Alarm

STEL

TWA

O

2

19.5%
Decreasing

23.5%

n/a

n/a


Combustibles

10% LEL

50% LEL

n/a

n/a

CO

25 ppm

50 ppm

200 ppm

25 ppm

H

2

S

5 ppm

30 ppm

5 ppm

1 ppm