GX-2009 Operator’s Manual

Calibration Mode • 38

6. Push the adapter plate onto the GX-2009’s sensor face. Make sure the

adapter plate is oriented as shown in Figure 12 below with the sensor
names on the adapter plate matching up with the sensor names on the
instrument.

Figure 12: Installing the Adapter Plate, One Cal

7. Press and release the POWER MODE button, the current gas reading

for the selected channel is displayed.

8. Turn the regulator on/off knob counterclockwise to open it. Calibration

gas will begin to flow.

9. Allow the gas to flow for two minutes.

10. Adjust the display gas reading using the AIR button to match the

calibration cylinder’s target gas concentration. See Step 5 on page 29
for instructions to adjust a parameter using the AIR button.

11. Press and release the POWER MODE button.