Metrohm 881 Compact IC pro – Cation User Manual

Page 19

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

881 Compact IC pro – Cation

■■■■■■■■

11

3 Installing the eluent path

Lead the eluent aspiration tubing (3-1) out of the instrument
through a capillary feed-through and connect it with the eluent
bottle (see Chapter 3.8.1, page 22).

Connect the (3-2) column inlet capillary and the detector inlet
capillary (3-3) with one another using a coupling (6.2744.040)
and two short pressure screws (6.2744.070).

4 Connecting the sample path

Guide the sample aspiration capillary (3-5) connected to the sam-
ple input of the injection valve out of the instrument through a
capillary feed-through and connect it with the Sample Processor,
if applicable (see Sample Processor manual).

Guide the sample outlet capillary (3-6) connected to the sample
output of the injection valve out of the instrument through a
capillary feed-through and onward to the waste container and
then fasten it there.

5 Connecting the instrument

Connect the instrument to a computer on which the software
MagIC Net™ is installed using the USB cable (6.2151.020) (see
Chapter 3.15.1, page 41)
.

Connect the instrument to the mains supply (see Chapter 3.15.2,
page 42)
.

6 Initial start-up

(see Chapter 4.1, page 47)

Switch on the PC and start MagIC Net™.

Switch on the instrument.

Deaerate the high pressure pump.

Rinse instrument without column.

7 Installing guard and separation column

Remove the coupling (6.2744.040) between the column inlet
capillary and the detector inlet capillary.

(Optional) Connect guard column (see Chapter 3.16, page 42)

– Fasten the guard column to the end of the column inlet

capillary (see leaflet to the guard column).

– Rinse guard column.