5 Operation and maintenance

881 Compact IC pro – Cation

■■■■■■■■

55

Cleaning/replacing the zirconium oxide piston

Clean one piston after the other as follows:

1 Removing the piston cartridge from the pump head

Loosen the piston cartridge with a wrench and unscrew from the
pump head by hand.

2

1

Figure 23

Pump head – removing the piston

1

Pump head

2

Piston cartridge

2 Dismantling the piston

Caution

On the inside of the piston cartridge there is a taut spring than can
jump out of the piston cartridge if suddenly loosing tension.

When opening the piston cartridge, hold pressure towards the
spring and unscrew carefully.

Loosen the screw of the piston cartridge with a wrench and
unscrew carefully by hand and by holding pressure towards the
taut spring.

Remove the zirconium oxide piston and lay on a tissue.

Remove the spring retainer, spring and the inner plastic sleeve
from the piston cartridge and lay by.