7.10 Mains connection
■■■■■■■■■■■■■■■■■■■■■■
76
■■■■■■■■
881 Compact IC pro – Cation
7.10
Mains connection
Required voltage
100…240 V ± 10 % (auto-sensing)
Required fre-
quency
50…60 Hz ± 3 (auto-sensing)
Power consump-
tion
■
65 W for typical analysis application
■
25 W standby (conductivity detector to 40 °C)
Power supply unit
■
Up to 300 W maximum, electronically monitored
■
internal fuse 3.15 A
7.11
Interfaces
USB
Input
1 USB upstream, type B (for connection to the PC)
Output
2 USB downstream, type A
MSB
2 MSB MiniDin 8-pin (female) (for Dosino, stirrer, remote lines, …)
Caution
When connecting an instrument to the MSB connector you must
switch off the 881 Compact IC pro.
Detector
1 DSUB-15-pin high density (female)
Column recogni-
tion
for an intelligent column
Leak sensor
1 jack plug
Further connec-
tions
Auxiliary
1 DSUB 15-pin (female)
Service