3.10 High pressure pump

■■■■■■■■■■■■■■■■■■■■■■

30

■■■■■■■■

881 Compact IC pro – Cation

3

Clamping screw

4

Eluent aspiration tubing
Eluent aspiration tubing (6.1834.080) or
(6.1834.090).

5

Backup ring

1 Connecting coupling

Fasten the coupling (15-2) with a pressure screw (15-1) on the pump
head inlet capillary (14-7).

2 Connecting eluent aspiration tubing

Caution

The clamping screws must be tightened carefully. To tighten, grip
the coupling (15-2) with the key (6.2739.000) and grip the clamp-
ing screw (15-3) with the wrench (6.2621.050).

Plug the eluent aspiration tubing (15-4) into the coupling (15-2).

Tighten the clamping screw (15-3).

3.10.2

Deaerating the high pressure pump

The high pressure pump will only operate perfectly if the pump head con-
tains no more air bubbles. Therefore it must be deaerated during initial
start-up and after every change of eluent.

Caution

The high pressure pump must not be deaerated before the initial start-
up .

Deaerate the high pressure pump as follows (see Figure 16, page 31):

Deaerate the high pressure pump

The instrument must be connected to the PC and switched on to deaerate
the high pressure pump.