background image

3.9 Eluent degasser

■■■■■■■■■■■■■■■■■■■■■■

26

■■■■■■■■

881 Compact IC pro – Cation

3

Wadding

4

CO

2

adsorber

Adsorbs CO

2

from the air (e.g. Merck soda

lime with indicator, no. 6839.10).

5

Eluent

6

Aspiration filter 6.2821.090

7

Filter holder
From accessories set 6.2744.210.

8

Clamping screw
From accessories set 6.2744.210.

9

Tubing weighting
From accessories set 6.2744.210.

10 Eluent bottle 6.1608.070

11 Bottle attachment 6.1602.160

12 SGJ clip 6.2023.020

13 Tubing nipple

14 Threaded stopper

3.9

Eluent degasser

Gas bubbles in the eluent lead to an unstable baseline, as high pressure
pumps can transport liquids, but not gases. The eluent therefore has to be
degassed, before it reaches the high pressure pump.

The eluent degasser removes gas bubbles and dissolved gases from the
eluent. For degassing, the eluent flows into a vacuum chamber through a
special fluoropolymer capillary.

Note

The eluent degasser is already installed in the newly delivered instru-
ment. The following installation instructions must only be followed, if
the connections to the degasser had to be disconnected for mainte-
nance.