9 injection valve, 1 protection, Injection valve – Metrohm 881 Compact IC pro – Cation User Manual

Page 75: Protection

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation and maintenance

881 Compact IC pro – Cation

■■■■■■■■

67

The time measured corresponds to the "transfer time". The rinsing
time should be at least 3 times the transfer time.

Checking the rinsing time

It is possible to determine whether the rinsing time is adequate via a direct
measurement of the sample carry-over. Proceed as follows:

1 Preparing two samples

Sample A: A typical sample for the application.

Sample sample B: Ultrapure water.

2 Determining "Sample A"

Let "Sample A" pass through the sample path for the duration of the
rinsing time, then inject and measure.

3 Determining "Sample B"

Let "Sample B" pass through the sample path for the duration of the
rinsing time, then inject and measure.

4 Calculating the sample carry-over

The degree of the sample carry-over corresponds to the ratio of the
peak areas of the measurement for sample B to the measurement for
sample A. The lower the ratio, the lower the sample carry-over. This
ratio can be modified by varying the rinsing time – thus allowing the
rinsing time required for the application to be ascertained.

5.9

Injection valve

5.9.1

Protection

To prevent contamination of the injection valve, a 6.2821.120 inline filter
(see Chapter 3.11, page 32) should be mounted between the high pres-
sure pump and the pulsation damper.

Advertising