3 drainage tubings, Drainage tubings, Figure 5 – Metrohm 881 Compact IC pro – Cation User Manual

Page 26: Leak sensor – plugging in

Advertising
background image

3.6 Rear of the instrument

■■■■■■■■■■■■■■■■■■■■■■

18

■■■■■■■■

881 Compact IC pro – Cation

Connecting the leak sensor

Connect the leak sensor as follows:

1 Pull out the leak sensor connection cable (5-3) from the base tray.

2 Plug the leak sensor connector plug (5-2) into the leak sensor con-

nection socket (5-1) on the rear of the instrument.

1 2

3

Figure 5

leak sensor – plugging in

1

Leak sensor connection socket
Is labeled with Leak Sensor.

2

Leak sensor connector plug

3

Leak sensor connection cable
Extractable. Coiled up in the base tray.

3.6.3

Drainage tubings

Fluid that escapes in the covering plate or in the detector chamber flows
through the drainage tubings into the base tray and past the leak sensor
into the waste container. This ensures that any leaks in the system will be
detected by the leak sensor.

Advertising