5.10 Separation column

■■■■■■■■■■■■■■■■■■■■■■

68

■■■■■■■■

881 Compact IC pro – Cation

5.10

Separation column

5.10.1

Separating efficiency

Which analysis quality can be attained, depends to a great extent on the
separating efficiency of the separation column used. The separating effi-
ciency of the selected separation column must be sufficient for the analy-
sis problems present. If difficulties occur, you should always first check the
quality of the separation column by recording a standard chromatogram.

You can find detailed information on the separation columns available
from Metrohm in the leaflet provided along with your separation column,
in the Metrohm IC-Column Program (available via your Metrohm
agent) or in the Internet at

http://www.metrohm.com

in the product area

Ion chromatography. You can request free information on special IC appli-
cations in the corresponding "Application Bulletins" or "Application
Notes
", which are available in the Internet at

http://www.metrohm.com

in the Applications area or via the Metrohm agent responsible.

5.10.2

Protection

To protect the separation column against foreign particles, which can
affect the separating efficiency, we recommend that both the eluent and
the samples undergo a microfiltration (filter 0.45 µm) before being aspira-
ted via the aspiration filter (6.2821.090).

We recommend always to use a guard column (see Chapter 3.16, page
42)
. This protects the separation column and considerably increases its
service life. Information regarding which guard column is suitable for your
separation column can be found in the Metrohm IC Column Program
(which is available from your Metrohm agent), the leaflet provided along
with your separation column, the product information on the separation
column at

http://www.metrohm.com

(product area Ion Chromatography)

or obtained directly from your agent.

The pulsation absorber (see Chapter 3.12, page 33) must be installed in
order to protect the column material from pressure concussion caused by
injection.

5.10.3

Storage

Always store the separation columns sealed and filled according to the
data of the column manufacturer when not using them.