6.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

72

■■■■■■■■

881 Compact IC pro – Cation

Problem

Cause

Remedy

Separation column – dead
volume at column head.

Install the separation column in reverse
flow direction (if permitted by the leaflet)
and rinse into a beaker.

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 45)
.

Precision problems -
significant scatter-
ing of the measured
values

Injection valve – sample
loop.

Check installation of the sample loop (see
Chapter 3.13.1, page 34)
.

Sample – rinsing volume
too low.

Increase rinsing time (see Chapter 5.8, page
66)
.

Injection valve – defective.

Request Metrohm Service.

Unexpected change
to the retention
times in the chroma-
tograms

Separation column –
diminished separating effi-
ciency.

Regenerate separation column (see Chap-
ter 5.10.4, page 69)
.

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 45)
.

Eluent – Gas bubbles in the
eluent.

Check connections of the eluent degasser (see
Chapter 3.9, page 26)
.

High pressure pump –
defective.

Request Metrohm Service.

Vacuum is not being
built

Eluent Degasser – Connec-
tor Vacuum on the rear of
the instrument not (tightly)
sealed.

Seal the connector Vacuum tightly with a
6.1446.040 threaded stopper.