Metrohm 858 Professional Sample Processor User Manual

Page 34

Advertising
background image

3.8 Mounting the sample tube

■■■■■■■■■■■■■■■■■■■■■■

26

■■■■■■■■

858 Professional Sample Processor

3

Nut
4.766.4320 (or 6.2833.020 + 6.2744.080)

4

Ring wedge

5

Needle holder

6

Sample tube
6.2846.000 (zirconium oxide) or
6.1835.020 / 6.1835.040 / 6.1835.050
(made of PEEK)

This is how you mount needle and capillary:

1 Remove the PEEK pressure screw

Loosen and remove the nut (23-3) screwed onto the needle holder.

2 Insert the needle

Insert the needle from above part way into the opening of the
needle holder (23-5).

Pull the PEEK ring wedge (23-4) down over the needle from
above. The narrow side of the seal must face upwards.

3 Fasten the needle

Screw the nut (23-3) into the needle holder. Slight pressure will
be required from below to push the needle upward while doing
this.

Use your hand to screw the nut in the needle holder tight (do not
use any tools!).

4 Connect the capillary

Pull the 6.2744.010 PEEK pressure screw (23-2) over the end of
the capillary.

Manually screw tight the PEEK pressure screw with the capillary in
the nut of the needle holder. The capillary must be pressed tight
while doing so.