Figure 32, Inserting the tubing cartridge – Metrohm 858 Professional Sample Processor User Manual

Page 40

Advertising
background image

3.12 Installing the peristaltic pump

■■■■■■■■■■■■■■■■■■■■■■

32

■■■■■■■■

858 Professional Sample Processor

Figure 32

Inserting the tubing cartridge

6 Connect the capillaries

Screw the respective capillaries tightly to the two tubing olives
with PEEK pressure screws (31-1).

Setting the flow rate
The contact pressure of the tubing cartridge must be adjusted in order to
regulate the flow rate. Proceed as follows:

1 Set the contact pressure

Press the contact pressure lever (31-5) all the way forward.

Switch on the drive of the peristaltic pump.

Then release the contact pressure lever gradually until the liquid is
suctioned for the first time.

Now release the contact pressure lever by two latch positions.

The contact pressure is now set optimally.