Metrohm 858 Professional Sample Processor User Manual

Page 39

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

858 Professional Sample Processor

■■■■■■■■

31

Mount the pump tubing as follows:

1 Remove the tubing cartridge

Release the tubing cartridge from the cartridge holder by pressing
the snap-action lever and unhooking from the mounting bolts (see
Figure 7, page 14)
.

2 Insert the pump tubing

Press the contact pressure lever all the way down.

Place the pump tubing in the tubing cartridge. The stoppers
(31-3) must snap into the corresponding holder of the tubing car-
tridge.

3 Connect the aspiration side

Place a 6.2744.030 tubing olive (31-2) on the aspiration side of the
pump tubing.

4 Connect the pressure side

Slide the union nut (31-6) of the 6.2744.160 pump tubing con-
nection (without filter) onto the pump tubing.

Select a suitable adapter (31-7) (depends on the outer diameter
of the pump tubing) and slide it onto the pump tubing.

Place the tubing olive (31-8) onto the pump tubing.

Screw the union nut (31-6) tight on the tubing olive (31-8).

5 Insert the tubing cartridge

Hang the tubing cartridge in the mounting bolt and press it
underneath in the cartridge holder until the snap-action lever
snaps in.