1 connecting dosing devices, Connecting dosing devices – Metrohm 858 Professional Sample Processor User Manual

Page 44

Advertising
background image

3.14 Connecting MSB devices

■■■■■■■■■■■■■■■■■■■■■■

36

■■■■■■■■

858 Professional Sample Processor

Note

When connecting MSB devices together, the following must be
observed:

Only one device of the same type can be used at a single MSB con-
nector at one time.

Type 700 Dosino and 685 Dosimat dosing devices cannot be con-
nected together with other MSB instruments on a shared connector.
These dosing devices must be connected separately.

Only one device of the same type can be used at a single MSB con-
nector at one time.

MagIC Net software supports only dosing devices of the 800 Dosino
type.

Caution

Exit the control software before you plug MSB instruments in. The con-
trol instrument recognizes when it is switched on which instrument is
connected at which MSB connector. The operating unit or the control
software enters the connected MSB devices into the system configura-
tion (Device manager).

MSB connections can be extended with the 6.2151.010 cable. The length
of the connection must not exceed a maximum of 15 m.

3.14.1

Connecting dosing devices

Three dosing devices can be connected to the instrument.

MagIC Net software supports only the 800 Dosino as a dosing device.

The types of dosing devices that are supported are:

800 Dosino

700 Dosino

805 Dosimat

685 Dosimat

Advertising