5 Troubleshooting

858 Professional Sample Processor

■■■■■■■■

45

5 Troubleshooting

5.1

Problems and their solutions

Problem

Cause

Remedy

Marked rise in pres-
sure

Injection valve – valve
blocked.

Rinse the valve with water in the opposite
direction or have it cleaned (by Metrohm serv-
ice technician).

The Swing Head
either misses the
rack positions
totally or is inaccu-
rate

Sample Processor – The
Swing Head is not correctly
configured.

Enter the correct values for Swing radius,
Swing offset etc. in the control software
under "Configuration".

Sample Processor – The
axial distance is not cor-
rectly configured.

Enter the correct value for Axial distance in
the control software under "Configuration".

Sample Processor – The
wrong rack table is being
used.

Initialize the rack using the function Initialize
rack
in the "Manual Control" of the control
software.

Peristaltic pump –
insufficient or no
delivery rate

Peristaltic pump – contact
pressure too weak.

Correctly set the contact pressure .

Peristaltic pump – filter
blocked.

Replace the filter .

Peristaltic pump – pump
tubing defective.

Replace the pump tubing .

Precision problems -
significant scatter-
ing of the measured
values

Injection valve – sample
loop.

Check the installation of the sample loop.

Injection valve – defective.

Contact the Metrohm Service.