2 connecting a stirrer or titration stand, Connecting a stirrer or titration stand, Figure 35 – Metrohm 858 Professional Sample Processor User Manual

Page 45: Connecting a dosing device

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

858 Professional Sample Processor

■■■■■■■■

37

Warning

If a Dosino is connected to the 858 Professional Sample Processor then
the connection cable must be equipped with a T.2400.102 ferrite core.
The ferrite core reduces any interference voltages that may occur and
thus ensures compliance with strict EMC standards pursuant to applica-
ble technical norms, see Chapter "Technical Data".

Proceed as follows:

1 Mounting ferrite core

Fasten a T.2400.102 ferrite core to the Dosino connection cable near
to the plug.

2 Connect a dosing device

Exit the control software.

Connect the connection cable to one of the sockets marked with
MSB on the rear of the control instrument.

Start the control software.

USB 1

Contr.

MSB 2

MSB 3

T.2400.102

Figure 35

Connecting a dosing device

3.14.2

Connecting a stirrer or titration stand

You can use a magnetic stirrer 801 Stirrer or 803 Ti Stand (stirring "from
below") or the 804 Ti Stand with a rod stirrer 802 Stirrer (stirring "from
above").

Connect a stirrer or a titration stand as follows:

1 Connect a stirrer or titration stand

Exit the control software.