TO R Q U E M O D U L E I N S T A L L A T I O N

Load Cell Calibration

Version 2

Model 224 Pit Eddy Current Brake Installation and User Guide

2-11

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

LOAD CELL CALIBRATION

This section provides instructions for using the Torque Module with WinPEP 7. The
Torque Module must be calibrated prior to use. Follow the directions on the screen
exactly. Failure to perform the directions accurately will result in improper torque
values.
1

Verify you are in the MakeRun screen.

2

Verify you are connected to the dyno electronics.
Note: For more information on connecting to the dyno electronics, refer to the
WinPEP 7 User Guide (on your WinPEP CD or at
www.dynojet.com/manuals.shtml) or the WinPEP 7 Online Help.

3

Select Tools

MakeRun Options

Torque Cell Calibration.

Note: Before proceeding, be sure the eddy current brake is free and clear of any
obstructions. There should not be anything resting on the eddy current brake or
the dynamometer drum during this procedure.

4

Click Next to perform the Zero Calibration.
The Calibration window will appear. The hardware is now zeroing out the torque
cell. If the unit does not calibrate, recheck the setup and retry.

Figure 2-11: Zero Calibration Window