Model 224 Pit Eddy Current Brake Installation and User Guide

C H A P T E R 2
Load Cell Calibration

2-16

While installing the calibration weights, you should notice the Torque Gauge on the
DynoTrac Window moving from 0 to about 500 foot-pounds.

Note: The Torque Gauge may or may not be in this range.
• If the torque cell has been previously calibrated incorrectly or has not been

calibrated for a while, the gauge may show values out of this range until

calibration is complete.

• If you use the bolt pattern in the center of the calibration arm, the gauge will

show values around 250-260 foot-pounds.

Note: Let the torque gauge needle stabilize before clicking Next.

9

From the Span Calibration window (Figure 2-14), click Next to continue.
At this point, the value on the gauge should match the value on the calibration
arm.

10 Remove the calibration arm and weights and click Finish.