Model 224 Pit Eddy Current Brake Installation and User Guide
C H A P T E R 2
Load Cell Calibration
2-12
Once the Zero Calibration is complete, the Calibration Mass window will appear.
5
Enter the Torque Module calibration value. Refer to Figure 2-13.
Note: You must perform this step the first time you calibrate the load cell.
Or
If you are only performing a Zero Calibration, click Finish.
Figure 2-12: Calibration Mass Window
Enter the calibration number stamped near the bolt pattern at the end of the
calibration arm. If you do not have enough room to use the bolt pattern closest to
the end of the calibration arm, use the number stamped near the bolt pattern in
the center of the arm.
Note: Dynojet recommends you secure the calibration arm using the bolt pattern
closest to the end of the arm unless space constraints in your dyno room do not
allow you to.
Figure 2-13: Calibration Arm Number
bolt pattern closest to
the end of the
calibration arm
bolt pattern near the
center of the
calibration arm