320IS Installation Manual

5.1.2

Reporting Commands

Reporting commands (see Table 5-2) send specific information to the EDP port. These commands can be used in

both setup mode and normal mode.

5.1.3

The RESETCONFIGURATION Command

The RESETCONFIGURATION command can be used to restore all configuration parameters to their default

values. Before issuing this command, the indicator must be placed in test mode (press and hold setup switch until

TEST

is displayed).

This command is equivalent to using the DEFLT function on the TEST menu. See Section 7.8 on page 55 for more

information about test mode and using the TEST menu.

All load cell calibration settings are lost when the RESETCONFIGURATION command is run.

5.1.4Parameter Setting Commands

Parameter setting commands allow you to display or change the current value for a particular configuration

parameter See Tables 5-3 through 5-12 for more information on menu parameters.
Current configuration parameter settings can be displayed in either setup mode or normal mode using the following

syntax:

command<ENTER>

Most parameter values can be changed in setup mode only. Use the following command syntax when changing

parameter values:

command=value<ENTER>

where

value

is a number or a parameter value. Use no spaces before or after the equal (=) sign. If you type an

incorrect command or value, the display reads

??

. Changes to the parameters are saved as they are entered but

typically do not take effect until you exit setup mode.
For example, to set the motion band parameter to 5, type the following:

MOTBAND=5D<ENTER>

Command

Function

DUMPALL

List all parameter values

IOVERSION

Display

I/O Module

software version

P

Display current displayed weight with units
identifier. See Section 7.2.1 on page 47 for
more information.

ZZ

Display current weight and annunciator
status. See Section 7.2.2 on page 47 for
more information.

S

Display one frame of stream format

Table 5-2. EDP Reporting Commands

Command

Description

Values

GRADS

Graduations

1–100 000

ZTRKBND

Zero track band

OFF, 0.5D, 1D, 3D

ZRANGE

Zero range

1.9%, 100%

MOTBAND

Motion band

1D, 2D, 3D, 5D, 10D, 20D, OFF

OVRLOAD

Overload

FS+2%, FS+1D, FS+9D, FS

DIGFLTRX*

Digital filtering

1, 2, 4, 8, 16, 32, 64

DFSENS

Digital filter cutout sensitivity

2OUT, 4OUT, 8OUT, 16OUT, 32OUT, 64OUT, 128OUT

DFTHRH

Digital filter cutout threshold

NONE, 2DD, 5DD, 10DD, 20DD, 50DD, 100DD, 200DD, 250DD

TAREFN

Tare function

BOTH, NOTARE, PBTARE, KEYED

SMPRAT

Sample rate

15Hz, 30Hz, 60Hz, 7.5Hz

NOTE: * X = 1, 2, 3

Table 5-3. CONFIG EDP Commands