320IS Plus Installation Manual - Configuration

31

3.2.6

Print Format Menu

See Section 6.0 on page 53 for information about custom print formatting. The PFORMT menu is used only if

the

320IS Plus

is used with the I/O Module option.

CONSTU

000000
number

Consecutive number start up value in the range of 000000–999999. Specifies the initial
consecutive number (CONSNU) value used when the indicator is powered on.

UID

000000
number

Specifies a unit identifier for the indicator in the range of 000000–999999. The unit ID can be
added to print ticket formats to identify the indicator used to generate ticket formats.

ACCUM

OFF
On

Accumulator — Specifies whether the accumulator is enabled. if enabled, accumulation
occurs whenever a print operation is performed.

DATE

DATFMT
DATSEP

Allows selection of date format and date separator. See level three parameter for descriptions.

TIME

TIMFMT
TIMSEP

Allows selection of time format and time separator. See level three parameter for descriptions.

SPPWD

0
number

Specifies a password access to setpoints while in weigh mode. If set to 0 (off), no password is
required for access to setpoints. Specifying a number configures password.

To make changes in weigh mode, press and hold UNITS key for 5 seconds. Display shows
SETPNT. Press GROSS/NET key. Display shows SPPWD. Press GROSS/NET key and enter
password. If no entries are made within 10 seconds, timeout occurs and indicator returns to
weigh mode. Only setpoints with ACCESS set to ON can be accessed in weigh mode.

STNDBY

000
0-255

Standby mode delay — Specifies the number of minutes the indicator must be inactive before
entering standby mode. Valid values are 000 (off) or 0-255 minutes. After typing a standby time
value, press the TARE key to enter this value. The display will go into standby mode, confirmed
by a blinking LED on the right side of the display. When in standby mode, power is still supplied
to the CPU and draws 1/2 of the current as when the display is powered. Press any key to exit
standby mode and reactivate the display.

The indicator enters standby mode if no key presses, serial communications, or scale motion
occur for the length of time specified on this parameter. Set this parameter to 000 to disable
standby mode.

Level 3 submenus

ACCESS

DSABLE
ENABLE

Operator access to piece count mode. Specify DSABLE if piece count mode will not be used.
With access disabled, pressing the GROSS/NET (MODE) key toggles between gross and net
modes only.

SPLSIZ

10
20
50
100
PCWGT
5

Sample size — Specify the default size used for counting scale operations. Sample size can be
changed in counting mode during sample acquisition.

DATFMT

MMDDYY
DDMMYY
YYMMDD

Specifies the format used to display or print the date.

DATSEP

SLASH
DASH
SEMI

Specifies the date separator character.

TIMFMT

24HOUR
12HOUR

Specifies the format used to display or print the time.

TIMSEP

COLON
COMMA

Specifies the time separator character.

PROGRM Menu

Parameter

Choices

Description

Table 3-6. Program Menu Parameters (Continued)