4 serial scale interface, 5 local/remote operation – Rice Lake 920i USB Installation Manual V5.05 User Manual

Page 123

Advertising
background image

Appendix

117

11.4

Serial Scale Interface

Serial ports 3 through 32 can be configured for serial
scale input. The serial scale function allows other scale
indicators to send gross, net, or tare weight data to the
920i. Once a serial port has been configured to accept
scale data, the data format can be customized to match
the data stream sent by that indicator.

To configure a serial scale, do the following:

1. Under the SERIAL menu, set the INPUT

parameter for the selected port to SCALE
( l e g a l - f o r- t r a d e s e r i a l s c a l e ) o r I N D U S T
(industrial serial scale).

2. Return to the SCALES menu. Under CONFIG,

drop down and select the serial port. If the serial
scale is not shown, press the

Change Type

softkey

to select available serial scales, then use the
navigational keys to select the serial scale. Press

Add

to move the scale to the righthand column,

the press

Done

.

3. Under the SERIAL menu, return to the selected

p o r t a n d s e t t h e f o r m a t u n d e r t h e S F M T
parameter to match the format sent by the serial
scale.

The default serial scale format is:

<2><P><W7.><U><M><S><CR><LF>

where:

<2>

STX character

<P>

Polarity

<W7.>

Seven characters of net data with decimal point

<M>

Mode

<U>

Units

<S>

Status

<CR>

Carriage return

<LF>

Line feed

NOTE:

Industrial serial scales (INDUST) do not require

the

<M>

,

<U>

, and

<S>

identifiers. However, the units and

number of decimal places must be specified. Units can be
selected from the FORMAT menu; decimal places should
be indicated on the w-spec identifier. For example, a
seven-digit weight requiring two decimal places should be
specified as

<W7.2>

rather than

<W7.>

.

See Section 11.6 for more information about stream
formatting and format identifiers.

iRev 4 provides several preset scale formats within its
Stream Formatting function. Figure 11-1 shows one of
the iRev 4 stream formatting displays.

Figure 11-1. iRev 4 Stream Formatting Display

11.5

Local/Remote Operation

For truck scale and similar applications, local/remote
support provides function equivalent to that of a
legal-for-trade remote display with keypad. Scale data from
the local indicator is also displayed at the remote unit, and
keypad input from the remote allows transactions to be
initiated from either the local or remote unit.

To configure for local/remote operation, first set up the
local scale (including softkey assignments, truck mode,
and database information, as required). Use the
SERIAL menu, serial commands, or iRev 4 to set the
Local Unit serial parameters shown in Table 11-7.
Configure the remote indicator using the serial
parameters listed for the Remote Unit.

Table 11-7. Local/Remote Configuration Parameters

Serial

Configuration

Parameter

Parameter Value

Local Unit

Remote Unit

EDP.INPUT#p

CMD

DISPLAY

EDP.STREAM#p

DISPLAY

KEYPAD

EDP.BAUD#p

115200 preferred; local and

remote values must match

EDP.ECHO#p

OFF

OFF

EDP.RESPONSE#p

ON

ON

In the listed serial commands, p represents the serial port
number.