iRev Calibration

59

4. Remove all weights from the scale. If chains or

hooks are used, place them on the scale. Press
the

Calibrate Zero

button to perform the zero

calibration (see Figure 4-6). A message box
appears when the process is complete.

Figure 4-6. iRev 4 Zero Calibration Display

5. Apply the test weights to the scale. Press the

Calibrate Span

button to perform the span

calibration (see Figure 4-7). A message box
appears when the process is complete.

Figure 4-7. iRev 4 Span Calibration Display

6. If the option for chains or hooks was selected

in step 3, the Rezero display is shown (see
Figure 4-8). Remove all weights from the
scale, including chains or hooks. Press the

Re-Zero

button to calibrate the zero offset.

Figure 4-8. iRev 4 Rezero Display

7. If performing a multi-point linear calibration,

up to five other calibration weight values can
be entered on the display shown in Figure 4-9.
The weights must be in ascending order and
must not include zero or the span weight. Enter
the weight values and click the

Go

button to

calibrate each point.

Figure 4-9. iRev 4 Linear Calibration Display

8. Review the new calibration values then click

Finish

to close the Calibration Wizard. To

restore the current calibration values, click

Cancel

.

Figure 4-10. iRev 4 Calibration Values Display