Calibrate

WZERO

WVAL

REZERO

1291

Deadload Calibration

+/-

Millivolts

Zero Count =>

Last Zero

Temp Zero

CALIBR

SCALE1

SCALES

iRev Calibration

57

Figure 4-2. WZERO Calibration Display

To calibrate the indicator using the front panel:

1. Place the indicator in setup mode (display

reads

Scale Configuration

) and remove all weight

from the scale platform. If your test weights
require hooks or chains, place the hooks or
chains on the scale for zero calibration.

2. With the

SCALES

menu highlighted, press the

down

key, then select the scale to be calibrated.

P r e s s

d o w n

a g a i n (

G R A D S

p a r a m e t e r

highlighted), then press

left

to highlight the

CALIBR

submenu (see Figure 4-1). Press

down

t o g o t o z e r o c a l i b r a t i o n (

W Z E R O

) . S e e

Figure 4-2 on page 57.

3. Ensure scale is empty, then press

down

again to

show the current WZERO value, then press the

Calibrate

softkey to calibrate zero. When

complete, the new A/D count for the zero
calibration is displayed. Press

enter

to save the

zero calibration value and go to the next
prompt (

WVAL

).

4. With

WVAL

displayed, press

down

to show the

stored calibration weight value. Use the
numeric keypad to enter the actual value of the
calibration test weights, then press

enter

to

save the value and go to span calibration
(

WSPAN

).

5. Place test weights on scale. Press

down

again

to show the current WSPAN value, then press
the

Calibrate

softkey to calibrate span. When

complete, the new A/D count for the span
calibration is displayed. Press

enter

again to

save the span calibration value and go to the
next prompt (

WLIN

).

6. Five-point linearization (using the WLIN

parameter) provides increased scale accuracy
by calibrating the indicator at up to five
additional points between the zero and span
calibrations.

Linearization is optional: if you choose not to
perform linearization, skip the WLIN
parameter; if linearization values have
previously been entered, these values are reset
to zero during calibration. WZERO and
WSPAN must be calibrated before adding
linearization points; WLIN values must be less
than the WSPAN value and must not duplicate
WZERO or WSPAN.

To perform linearization, follow the procedure
below:

With

WLIN

displayed, Press

down

to go to the

first linearization point (

POINT 1

). Press

down

again to show the weight value prompt (

WGT

1

), then

down

once more to show the weight

value. Place test weights on the scale, then use
the numeric keypad to enter the actual test
weight value. Press

enter

to save the value and

move to the calibration (

CAL 1

) prompt. Press

down

to show the current calibration value,

then press the

Calibrate

softkey to calibrate the

linearization point. When complete, the A/D
count for the linear calibration is displayed.
Press

enter

again to save the calibration value

and go to the next prompt (

POINT 2

).

Repeat for up to five linearization points. To
exit the linearization parameters, press the

up

key to return to WLIN.

7. The optional rezero function is used to remove

a calibration offset when hooks or chains are
used to hang the test weights.

Note

The rezero function cannot be used with
five-point linear calibration.

If no other apparatus was used to hang the
test weights during calibration, remove the
test weights and press

up

to return to the

CALIBR submenu.

If hooks or chains were used during
calibration, remove these and the test
weights from the scale. With all weight
removed, go to the REZERO parameter,
then press

down

to show the current zero

value. Press the

Calibrate

softkey to adjust

the zero and span calibration values. Press

e n t e r

or

u p

to return to the CALIBR

submenu.

8. Press

up

to return to the SCALES menu, or

press the

Save and Exit

softkey to exit setup

mode.