Appendix

123

11.8

Data Formats

Continuous Output Serial Data Format

If continuous transmission is configured for a serial
port (STREAM parameter set to LFT or INDUST on
the SERIAL menu), the 920i sends data using the
Consolidated Controls serial data format shown in
Figure 11-2:

<STX> <POL> <wwwwwww> <UNIT> <G/N> <S> <TERM>

ASCII 02
decimal

Polarity:
<space> = Positive
<–> = Negative

Weight: 7 digits, right-justified, dummy
zeroes, decimal point with no leading
zeroes except for leading zero immediately
preceding the decimal point. Leading
zeroes transmitted as spaces.

L = pounds
K = kilograms
T = tons
G = grains
<space> = grams
O = ounces

G = Gross

N = Net

Status:
<space> = valid
I = Invalid
M = Motion
O = Over/under range
Z = Center of zero

<CR> <LF>

or <CR>

Figure 11-2. Continuous Output Serial Data Format

Demand Output Serial Data Format

When demand mode is configured for the serial port
(STREAM parameter set to OFF), the 920i uses a data
string formatted for a basic ticket printout. The
particular ticket format printed depends on the
indicator configuration.

You can customize the ticket to work with a wide
variety of printers, scoreboard displays, and other
remote equipment. See Section 7.0 on page 67 for
more information on custom print formats.

RS-485 Data Formats

Two-wire RS-485 communications is available on port
4 of the CPU board; four-wire RS-485 communications
is supported on the “A” ports of any installed serial
expansion cards.

The 920i has a built-in RS-485 software protocol
which is enabled when you assign a non-zero address
to the indicator. Valid RS-485 addresses must be in the
range 1–255; the address is specified on the ADDRESS
parameter on the SERIAL menu.

All remote commands are initiated using the data
format shown in Figure 11-3:

<STX> <ADDRESS> <COMMAND> <CR>

ASCII 02

decimal

Address of the

receiving indicator

EDP serial command

NOTE: Host must send <CR>, not

<CR><LF>. Failure to use <CR>
renders all indicators unable to
respond to serial commands.

ASCII

13 decimal

Figure 11-3. RS-485 Send Data Format

If the initiating device address matches the port address
of an 920i on the RS-485 network, that indicator
responds. For example, with demand outputs, or in
response to an XG#1 command, the responding
indicator uses the format shown in Figure 11-4:

<STX> <ADDRESS> <DEMAND DATA> <ETX> <CR>

ASCII 02

decimal

Address of the
transmitting indicator

Response commands from
indicator:
<first line> <EOL>
<additional lines> <EOL>
<last line> <EOL>
where EOL = <CR> or <CR> <LF>
(set by TERMIN parameter on the
SERIAL menu)

ASCII 13

decimal

ASCII 03

decimal

Figure 11-4. RS-485 Respond Data Format

Example:

To send the XG#1 command from an ASCII

terminal to an indicator at address 65 (decimal) on the
RS-485 network, use the format shown in Figure 11-3.

The keyboard equivalent for the start-of-text
(

STX

) character is

CONTROL-B

.

The indicator address (65) is represented by an
upper case “A”.

The carriage return (CR) character is generated
by pressing the

ENTER

key.

Therefore, to send the XG#1 command to the indicator
at address 65, enter the following at the terminal:
<CONTROL-B>AXG#1.

The indicator responds with the format shown in
Figure 11-4:

<STX>A 1234.00 lb<CR><LF><ETX><CR>