14 ScCMP Operator’s Manual

4.1.2

Re-Calibration

To perform a Re-Calibration:

1. Select the sensor (load cell) to calibrate from the left status panel.

2. Select

Re-Calibration

from the drop down box in the controls interface.

3. Press the

Next

button.

Zero Reference Point

4. Unload the scale (with the exception of any lift rigging required during calibration) to establish the zero

(unloaded) reference point.

5. Once the scale is unloaded and stable, press

Next

to continue.

Note

Note

Note

Note

The ADC Count displayed on the left status panel can be used as a reference to indicate the current load and
stability. This value will never be perfectly stable, in ideal situations expect the value to jump 50 counts or more.

Span Reference Point

6. Load the scale with a known test weight, plus the same lift rigging used when establishing the zero reference.

7. Enter the test weight value into the calibration dialog.

8. Once the scale is loaded and stable, press

Next

to continue.

As above, the ADC Count display can be used as a reference to indicate the current load and stability. This value
should be different from the zero reference ADC Count value. If not, ensure you have the correct sensor
selected for calibration.

The Cal Points tab provides a look at the current calibration table. Since there is only one point entered, all
points will be equal at this point.

Additional Span Reference Point(s)

If no additional span reference points are required, skip to step 12.

9. To establish additional span points, load the scale with the next test value.

10. Enter the test value into the calibration control dialog.

11. Press the

Next

button.

Save/Complete Calibration

12. To save and complete this calibration, press the

Save/Complete

button.

13. Wait for ScCMP to complete calibration (confirm by checking the Status value on the left).

Note

The Cal Points tab provides a look at the current calibration table. From this table you can see the ADC Count
and the calibration value that was captured for each reference point.

4.1.3

Cal-Number

To perform a Cal-Number:

1. Select the sensor (load cell) to calibrate from the left status panel.

2. Select

Cal-Number

from the drop down box in the controls interface

3. Press the

Next

button.

Zero Reference Point

4. Unload the scale to establish the zero (unloaded) reference point.

5. Once the scale is unloaded and stable, press

Next

to continue.

Note

Note

The ADC Count displayed on the left status panel can be used as a reference to indicate the current load and
stability. This value will never be perfectly stable, in ideal situations expect the value to jump 50 counts or more.

Enter Cal-Number Value

6. Enter the Cal-Number value in the calibration control dialog.

7. Press

Next

to complete the Cal-Number calibration.

The Cal-Number value may be found stamped on the load cell or recorded in previous calibration records.