Advanced Configuration

21

6.2 Sensors

The advanced sensor view provides a list of all sensor types and their associated configuration.

Figure 6-2. Configuration View - Sensors

6.2.1

Parameters

Table 6-2. Sensor Parameters

Parameter

Description

ID

Sensor ID index for every type of sensor. Typical range 0 to 16.

Enabled

Enable or Disable the sensor.

Type

Sensor type, dependent on the hardware configuration.

Name

Name of this sensor. Maximum number of characters is 20.

CnvSpeed

Sensor conversion speed range 1 to 15. Smaller value faster the speed and more noise. This value is
hardware dependent.

ACExc

AC Excitation feature enabled. This feature is not yet supported in many ScaleCore products.

TempComp

Temperature Compensation feature enabled. This feature is not yet supported in many ScaleCore products.

Vref

Voltage Reference enable. This feature is hardware dependent.

SWFilter

The software filter level for the sensor. Typical range 0-3 where the lower number is less filtering.

Motion Enabled

Enable or Disable motion detection.

AZM Enabled

Enable or Disable AZM (Auto Zero Maintenance) algorithm.

Zero at Power Up

Enable or Disable automatic zero of load at power up.

Standard Mode

Set the weighing mode (Industry, NTEP, OIML, 1-Unit).

6.2.2

Controls

Edit Sensor – This control allows editing the sensor configuration parameters.

Edit Standard Mode This control allows editing of the standard mode, including motion detection, AZM and
Zero at power up.