Calibration Tools

15

4.2 DAC Controls

ScCMP provides controls for DAC (Digital to Analog) output functions in a ScaleCore product. Controls include
calibration and manual control.

Note

DAC outputs may not be available in all ScaleCore family products.

4.2.1

Calibrate DAC

ScCMP provides an interface for performing calibration of the available Digital to Analog Converter (DAC)
outputs. The DAC calibration interface is divided into two sections.

• The left provides the DAC Status interface. This shows the currently selected DAC output.

- Current calibration values – Offset, Gain, Minimum and Maximum Span.
- DAC status – Calibrated, Enabled, In Cal, and Manual Mode.
- DAC configuration relative to input sensor – Sensor ID, Units, Value Type, Minimum and Maximum

Value.

• The right pane provides the calibration controls interface.

- At the top of this will be some brief instructions for what each calibration step requires.

Note

Please reference your specific hardware user guide for details on connecting to and measuring the DAC output.

To perform a DAC calibration:

1. Select the DAC output to calibrate from the left status panel.

2. Press the

Next

button.

Offset

The DAC voltage output will go to a mid-range value. To adjust the offset:

3. It should be approximately equal to the mid-range of the desired DAC output. If needed, adjust the output

value up or down with the calibration controls.

4. Press the

Next

button to continue.

Note

Typically this value is left at default (0).

Gain

The DAC voltage output will go to a full scale value. Adjust the gain:

5. If needed, adjust the output value up or down with the calibration controls.

6. Press the

Next

button to continue.

Note

Typically this value is left at default (0).

Minimum Span

The DAC voltage output will go to the absolute minimum value. Adjust the minimum span:

7. This value coincides with the minimum value assigned to the sensor configuration. If needed, adjust the

output value up or down with the calibration controls.

8. Press the

Next

button to continue.

Maximum Span

The DAC voltage output will go to a full scale value. Adjust the maximum span:

9. This value coincides with the maximum value assigned to the sensor configuration. If needed, adjust the

output value up or down with the calibration controls.

10. Press the

Next

button to complete the calibration.