1 parameters, 2 controls, Parameters – Rice Lake MSI-8000 RF Remote Display - ScaleCore Configuration Management Program (ScCMP) User Manual
Page 29: Controls
Advanced Configuration
25
6.4.1
Parameters
Table 6-4. Calibration Table Parameters
Parameter
Description
Table cal points
Cal Table
Calibration Table Number.
Sensor ID
Sensor ID associated with this cal table.
Status
Status of this cal table.
Capacity
Capacity of a sensor.
Count by
Countby of a sensor.
Units
Sensor value unit.
Temp Zone
Temperature zone for this calibration table (temperature compensated calibration not available in all
systems).
Table cal points
Cal Point
Cal point index.
ADC Count
ADC count.
Value
Value of sensor. Weight for load cell.
6.4.2
Controls
Edit Calibration Point – allows manually editing a single calibration point within a calibration table without
performing a calibration.
Important
Adjusting calibration table points will affect the current sensor calibration.
Edit Cal Table – allows editing of the ADC count and Values in a table by typing into a spreadsheet style interface.
Following diagram shows the Calibration Table Editor.
Figure 6-5. Calibration Table Editor
Additionally, a convenience function is provided to Zero the calibration table. This function accepts numeric input
as a value to be added to every ADC count in the editor.
This function is helpful when transferring calibration from one set of electronics to another and correcting for zero
offsets.