Setup/Configuration Mode Access


There are two ways to access the Setup/Configuration Mode:

1) Using the Calibration Jumper (JP1)

i) Turn off the instrument and insert Jumper 1 (JP1).

ii) Turn the instrument on, the display will show the Audit Trail then

prompt for password entry [P 0].

iii) Enter the password “2001” and press the [Enter] key.

2) Return Jumper 1 (JP1) to its original position when setup/calibration is

completed.

3) “Hot Key” entry if enabled (C1.2 is set to Enable.)

i) Press the [G/N] and [Enter] keys together while in normal operation

mode

ii) The display will show the Audit Trail then prompt for password entry

[P 0].

iii) Enter the setup/calibration password and press the [Enter] key.

Configuration/Calibration Mode

Category Selection:

The setup mode is arranged into major categorie s and their individual

selections. Upon entering the setup mode the display will prompt ‘C1’

indicating category ‘1’. Using the ‘Print’ and ‘G/N’ keys scroll through the setup

categories until the desired category is displayed. When the desired categ ory

is displayed press the ‘Enter’ key to begin setup. Pressing ‘Zero’ while in the

category menu will prompt ‘END’, pressing ‘Enter’ at this prompt exits the

setup menu.

Setup Selection:

After entering the desired category use the ‘Print’ and ‘G/N’ keys to toggle

setup choices or alter setup values. Press ‘Enter’ to accept when the desired

value is displayed to save the value and advance to the next selection.

Note: Features requiring the Group B Function keys will not be available

when configuring a standard

Series 9.