Rice Lake DWM-IV Weighmeter Series 9 Setup Manual User Manual

Page 18

Advertising
background image

16

Configuration

Step

Values

Set

Description

C6

Calibration

REF

Calibration Mode

RC

Standard Calibration (Zero Calibration and Span Calibration)

RS

Span Calibration Only

C6.1

RO

Zero Calibration Only

PCAL

Zero Calibration

Display will show PCAL about 1 sec then “R XXXX.XX

C6.2

Insure the scale is unloaded and press ‘Enter’

CAL

Span Calibration

Display will show “CAL” about 1 sec then “N XXXX.XX”

Key in weight value to be used for Span Calibration (Optional Series 9)

Scroll in the weight value to be used for Span Calibration (Std. Series 9)

Place the Weight on the scale and p ress ‘Enter’

C6.3

Press Enter key to save the Calibration when “End” is displayed

C7

Calibration Tune

C7.1

TUN

Tune Calibration (Adjust Span)

0~255

1 = 0.1 times the current value of the one division

C7.2

DIR

Tune Direction

UP

Adjust span upward

DOWN

Adjust spa n downward

(Optional Series 9 Unit Only) To Enter the Custom Print Format Procedure

With the C1 category displayed press ‘Shift’ then ‘0’

C10

Custom Print Formats (See details in the Customizing Print Format
section)

‘P1 CUST’

C10.1

Edit the Custom Print Format

‘P2 CONT’

C10.2

Edit the Custom Continuous Print Format

‘FMT END’

C10.3

Quit Print Format

Advertising