Calibration Audit Trail


The audit trail consists of two event counters, one for the adjustment
(calibration) parameters (audit trail 1) and one for configuration parameters
(audit trail 2). The range will be (000 to 999) for each counter.

To view the audit counters press the ‘G/N’ and ‘Enter’ keys at the same time
while the instrument is in the normal weighing mode. The audit trail will be
displayed as long as the keys are held.

If the calibration jumper is installed the audit trail will be displayed momentarily
before the password prompt.

The audit trail format will be:

AXXX.YYY where:

XXX is audit trail 1 and YYY is audit trail 2.


Audit trail one will increment when changes are made to:

Pre-calibration data
Calibration
Calibration Tune


Audit trail two will increment when changes are made to:

Filter settings

Print related settings C3.6-C3.9,
Print Format

Weight related operation keys such as Tare mode