Error messages, Jumper settings – Rice Lake DWM-IV Weighmeter Series 9 Setup Manual User Manual

Page 23

Advertising
background image

21

Error Messages


Error Messages Explanation

CAP OL

Over Capacity, weight value exceeds the full scale capacity

ADC UL

Input load cell signal is too low or reversed, load cell or load cell -cable is faulty

ADC OL

Input load cell signal is too high, load cell or load cell -cable is faulty

NOT CAL

Scale is Not yet Calibrated

RMT DSP

Will be displayed on Remote Display Unit when there is no signal from th e Main
Display

RMT OPE

Will be displayed on Remote Operation Unit when there is no signal from the Main
Display

SPAN OL

Maximum calculated span count is greater than m aximum ADC Chip count

LO BAT

Insufficient battery voltage level. Need to re -charge the battery

- - - - - - -

Display if Power On Zero fails. Need to unload the scale or optional press Zero
Key if power on zero weight is within allowable zero r ange

S.TOTAL

Displayed before entering the password if Sub Total printing is pending. Need to
press ‘Shift’+’Print’ to print the Sub Total and ‘Shift’+’G/N’ to print the Grand Total

G.TOT AL

Displayed before entering the password if Grand Total printing is pending. Need to
press ‘Shift’+’G/N’ to print the Grand Total

Jumper Settings

Reference Table For Jumpers

JP-1

Calibration Enable

JP-2

Reserved

JP-3

Not in use

JP-4

Not in use

JP-5

Not in use

JP-6

to set Excitation Voltage to 5V (Use to conserve power d uring battery operation)