Calibration and Adjustment

Purpose

Calibration is the determination of any

difference between the measured value

displayed and the true weight (mass) of

a sample. Adjustment is the correction

of this difference, or its reduction to

an allowable level within maximum

permissible error limits.

Features

Calibration | adjustment can be

performed only when

– there is no load on the balance

– the balance is tared

– the internal signal is stable

– the weight displayed for the sample on

the balance must not differ from the

nominal weight by more than 2%.

If these conditions are not met, an error

message is displayed “Err 02".

Following calibration | adjustment,

the application program is cleared.

Internal Calibration | Adjustment

Note:

Only for models with the label

Entris...i-1S !

In the operating menu, select

Cal.Just. - cal.int. before beginning.

The built-in motorized calibration

weight located in the housing is applied

and removed automatically for internal

calibration.

t

Select calibration | adjustment:

Press Q

> The internal calibration weight is

applied automatically

> The balance is adjusted | calibrated

> The built-in calibration weight is

removed