Rice Lake Entris, Sartorius Laboratory User Manual

Page 14

Advertising
background image

14

External Calibration

Set the following parameters:
SETUP - BAL.SCAL.- CAL.Just. - Cal.Ext. (Code 1.1. 9. 1)

The required calibration weight is configured at the factory (see “Specifications").

Step

Key (or instruction)

Display

1. Tare balance

w

0.0

g

2. Start adjustment routine

Q

Cal.Ext.

Once you store the zero point

- 5000.0

g

the display prompts for the required

calibration weight (flashing display)

3. Apply the prompted calibration

5000.0

g

weight (in this example 5000 g).

Weight too low:

a minus sign “-" is shown

Weight too high:

a plus sign “+" is shown

The display stops flashing as soon as

the weight value is within the defined

limit.

4. Calibration/adjustment executed;

Cal.end

then the calibration weight

is displayed

+ 5000.0

g

5. Remove the calibration weight

0.0

g