Troubleshooting guide – Rice Lake Entris, Sartorius Laboratory User Manual

Page 35

Advertising
background image

Error codes are shown on the main display for approx. 2 seconds.
The program then returns automatically to the previous mode.

Display Cause

Solution

No segments appear on

No AC power is available

Check the AC power supply

the display

The power supply is not

Plug in the power supply

plugged in

high

The load exceeds the

Unload the balance

balance capacity

low or Err 54

Something is touching the

Move the object that is touching

weighing pan

the weighing pan

App.err.

Cannot store data:

Increase load

Load on weighing pan too

light or no sample on pan

while application is active

dis.err.

Display error:

Change the configuration in the

Data output not compatible

operating menu

with output format

prt.err.

Interface port for printer

Reset the menu factory settings

output is blocked

or

Contact your local

Sartorius Service Center

err 02

Calibration parameter not met, Calibrate only when zero is displayed

e.g.:

– Press w to tare the balance

– load on weighing pan

Unload the balance

err 10 The

w

key is blocked

After the tare memory has been

for active application programs; deleted using the c key,

Only 1 tare function can be

the w key can be used again

used at a time

err 11

Tara memory not allowed

Press w

The weight readout

Unstable ambient conditions

Set up balance in another area

changes constantly

(excessive vibration or draft)

Adjust Setup configuration

A foreign object is caught

Remove the foreign object

between weighing pan

and housing

The weight readout

The balance was not

Calibrated/adjust the balance

is obviously wrong

calibrated/adjusted

Balance not tared before

weighing

If any other errors occur, contact your local Sartorius Service Center.
Web address:
http://www.sartorius.com

Troubleshooting Guide

35