Example: Counting parts of equal weight
Parameter settings: Applic. - Count. (menu code 2. 3.)
Step
Key (or instruction) Display/Data output
1. Place empty container
+
22.6
g
on the balance
2. Tare the balance
w
0.0
g
3. Add reference sample quantity to
container (in this example: 20 pcs)
4. Changing the reference sample quantity
b
ref
10
pcs
5. Select reference sample quantity:
Repeatedly:
b
In increments of 1 (1, 2, 3, ..., 100)
Press briefly
b
In increments of 10 (10, 20, ..., 100) press and hold
ref
20
pcs
6. Confirm selected reference
u
+
20
pcs
*
sample quantity and start
nRef 20 pcs
the application. The current
wRef 1.07 g
reference weight remains saved
until a new reference is set or
the power supply is interrupted.
7. Add desired number of pieces
+
500
pcs
8. If desired, print quantity
r
Qnt + 500 pcs
9. Toggle display between mean
+
1.07
g
G
*
piece weight, weight, quantity
+
535.0
g
*
Repeatedly:
b
+
500
pcs
*
10. Unload the balance
–
21
pcs
*
11. Repeat as needed, starting from Step 7
12. End “Counting"
c
0.0