Density determination – Rice Lake Entris, Sartorius Laboratory User Manual

Page 31

Advertising
background image

31

Density Determination

Display symbol: W

Purpose

This application program lets you determine the density of solid substances using the

buoyancy method.

Features

To enter the density of the buoyancy liquid(g/cm

3

) at the corresponding temperature,

press b. See the next page for a table of density values for water. The factory setting

is 1 g/cm

3

.

The following formula is applied:

Density of sample =

Weight in air

———————————————————— + density of liquid

(Weight in air – weight in water)

When you start the density determination routine, the density of the liquid is displayed

briefly.

Positive and negative values can be stored for weight in air and weight in water.

The weight in water must be less than the weight in air; otherwise, an error message is

displayed.

The results can be displayed with 0 to 3 decimal places: see “Configuration". Not part of

the scope of delivery: sample holder and suspension wire.

Preparation

t

Select the Density Determination application in the menu:

see “Configuration"

t

Set the following parameters:

Applic. ation programs
density determination
DEC.PLCS Decimal places

none No decimal places

o

1 dec.pl.

1 decimal place

2 dec.pl.

2 decimal places

3 dec.pl.

3 decimal places

o = Factory setting

Note on using 3 decimal places:

Using three decimal places for

density can result in a high

measurement error rate because

corrections to the air density and

the density calculation sets are not

taken into account, for example.

Advertising