Model-specific Specifications

Modele:

Entris

224-1S 124-1S 64-1S

224i-1S 124i-1S 64i-1S

Weighing capacity

220 g

120 g

60 g

Readability

0.0001 g

0.0001 g

0.0001 g

Tare range (subtractive)

220 g

120 g

60 g

Repeatability

< ±

0.0001 g

0.0001 g

0.0001 g

(standard deviation)
Linearity

0.0002 g

0.0002 g

0.0002 g

Response time (average)

s

2,5

2,5

2,5

Sensitivity drift within

<±/K

2 · 10

–6

2 · 10

–6

2 · 10

–6

+10 ... +30 °C
Adaptation to ambient conditions

By selection of 1 of 4 optimized filter levels;

display update: 0.1–0.4 (depends on filter level selected)

External calibration weight

g

200 (E2)

100 (E2)

50 (E2)

(of at least accuracy class...)
Net weight, approx.

kg

4.4 | 4.8

4.4 |4.8

4.4 | 4.8

Weighing pan size

mm

90 d 90

d

90

d

Whg. chamber height

mm

230

230

230

Dimensions (W + D + H)

mm

230 + 303 + 330

Modele:

Entris

623-1S 423-1S 323-1S

623i-1S 423i-1S 323i-1S

Weighing capacity

620 g

420 g

320 g

Readability

0.001 g

0.001 g

0.001 g

Tare range (subtractive)

620 g

420 g

320 g

Repeatability

< ±

0.001 g

0.001 g

0.001 g

(standard deviation)
Linearity

0.002 g

0.002 g

0.002 g

Response time (average)

s

1

1

1.1

Sensitivity drift within

<±/K

2 · 10

–6

2 · 10

–6

2 · 10

–6

+10 ... +30 °C
Adaptation to ambient conditions

By selection of 1 of 4 optimized filter levels;

display update: 0.1 – 0.4 (depending on the

set filter level)

External calibration weight

g

500 (F1)

200 (F1)

200 (F1)

(of at least accuracy class...)
Net weight, approx.

kg

3.2|3.6

3.2|3.6

3.2|3.6

Weighing pan size

mm

115 d 115

d

115

d

Dimensions (W + D + H)

mm

230 + 303 + 136