Parameter settings:
Applic. - Density - dec.plcs. -1 dec.pl. (menu code 2. 9. 1. 2)

Example: Determining the density of a solid using water as the buoyancy liquid.

The density of water at 20°C is 0.99823 g/cm

3

.

Step

Key (or instruction) Display/Data output

1. Attach sample holder and suspension wire
2. Tare the balance

w

0.0

g

3. Edit the stored density value

b

_1.00000

4. Enter the density of the liquid

Repeatedly: b,

(in this example: 0.99823): Enter

briefly or press and

numerals in increments of 1

hold ;

u

, etc.

_0.99823

5. Save density value and start

application. The density value

u

is stored in battery-backed memory

until the setting is changed.

6. Confirm “air" display

u

air

?

7. Determine the weight of

sample in the air:

Place sample on the balance

+

20.0

g

?

*

8. Store value for weight in air

u

9. Remove sample from the balance

water

?

10. Determine weight in liquid:

Place sample in holder.

11. Confirm “water" display

u

0.0

g

?

*

12. Immerse sample in liquid

+

15.0

g

?

*

13. Store value for weight in

u

liquid, view and print result

+

4.0

o

?

*

RhoFl 0.6237 o

Wa + 20.0 g

Wfl + 15.0 g

14. Delete result

c

Rho 4.0 o

15. Repeat as needed, starting from Step 5.