Rice Lake Entris, Sartorius Laboratory User Manual

Page 26

Advertising
background image

26

Example: Determining residual weight in percent

Parameter settings: Applic. - percent (menu code 2. 4.)

Reference percentage: Ref 100%

Step

Key (or instruction) Display/Data output

1. Tare the balance

w

0.0

g

2. Changing the reference:

(see the previous page)

b

ref 100

%

3. Place sample equal to 100% on the

balance (in this example: 111.6 g)

4. Start the application.

u

+

100.0

%

*

The current reference weight

pRef 100 %

remains stored until a new

Wxx% + 111.6 g

reference is set or power to the

power supply is interrupted.

5. Remove sample (e.g. for drying)

6. Place weight on the balance

+

94.9

%

*

(in this example 322.5 g)

7. If desired, print percentage

r

Prc + 94.9 %

8. Toggle display between

+

105.9

g

*

weight and percentage

Repeatedly:

b

+

94.9

%

*

9. Clear display of residual weight

c

+

105.9

g

and reference percentage

Exit application

10. If desired, print net residual weight

r

N + 105.9 g