Preparation

t

Select the Counting application in the

menu:

see “Configuration"

t

Set the following parameters:

Applic.ation programs
Count.
Resolut.ion
o

Disp.Acc. Display accuracy

10-fold 10-fold higher

Ref.Updt. Autom. ref. sample

updating

o Off Off

autom.

Automatic

o = Factory setting

Printout: Counting

nRef 10

: Ref. sample quantity

wRef 21.14 g

: Reference weight

for 1 unit

Qnt + 500 pcs

: Calculated quantity

23

Application Programs

Counting

Display symbol:

Z

Purpose

With the Counting application, you can

determine the number of parts that each

have approximately equal weight. To

do this, a known number of parts (the

reference sample quantity) is weighed first,

and the individual piece weight (reference

weight) is calculated from this result. Thus

the number of parts subsequently placed

on the balance can be determined from

their weight.

Changing the Reference Sample Quantity

Activate function:

Press the b key

Select the desired reference sample

quantity (1 to 100):

In increments of 1: Press the b key briefly

In increments of 10:

Press and hold the b key.

The quantity is stored in battery-backed

memory.

Reference Sample Updating

Automatic reference sample updating

optimizes the counting accuracy. You can

activate or deactivate this function in the

menu.

Automatic reference sample updating

is performed when the requirements,

including the specified stability criterion,

have been met.

The abbreviation opt for “optimizing”, is

displayed briefly with the new reference

sample quantity.