Levelling the Balance

Purpose:

– To compensate for unevenness at the place of

installation

Always level the balance again any time after it has

been moved to a different location. Only the 2 front

feet are used for leveling.

t

Screw in both rear support feet (only on models with

a rectangular weighing pan).

t

Turn the front levelingfeet as shown in the

illustration until the air bubble is centered within

the circle of the level indicator.

> Normally, several leveling steps are required.

t

On models with a rectangular weighing pan:

Screw out both back leveling feet until they touch

the setup surface.

Below-Cell Weighing

A port for a below-cell weighing hook is located on

the bottom of the balance.

y

Not permitted for applications in legal metrology.

t

Lift cover plate out of the bottom of the balance.

Attention: Place the balance on its side, do not turn

over completely!

t

Secure hook

1: Use a wire, for example, to suspend

the sample on the hook.

y

Install a draft shield if necessary.